Brand Name : JINHAIHU
Model Number : JCB-2
Certification : ISO9001
Place of Origin : BEIJING, CHINA
MOQ : 1 set
Payment Terms : T/T,L/C,Western Union,MoneyGram,D/A,D/P
Supply Ability : 20 sets per month
Delivery Time : 7-15 working days
Packaging Details : Standard export wooden cases
Place of origin : Beijing, China
Product Name : ISO1182 Building materials noncombustible test furnace
Model : JCB-2
Furnace specification : D: Φ40cm; High:1.5m
Sample specification : Φ45×50(mm)
Heating furnace pipe parameter : D: Φ75mm±1mm; De: Φ90mm±1mm; H:150mm±1mm
Environmental conditions : -5℃—40℃
Supply voltage : AC220V±10%, 50Hz,Current≤10A
Temperature in furnace : 750±5°C,drift ≤2℃ within 10min
Temperature measurement accuracy : ≤±0.5℃
Building materials noncombustible test furnace
Scope of application:
Product function characteristics:
1.Utilizes microcomputer and silicon controlled technologies to adjust furnace body temperature via power regulation.
2.Dual control mode: Operable via touch screen or computer, with easy mode switching; One device supports multiple computers simultaneously for LAN sharing (user provides computer and printer; host computer software, router, and network cable included with equipment).
3.High automation: Displays and records furnace surface temperature, sample center temperature, and real-time data during testing, with automatic operation prompts for each test step.
4.Queryable historical curves for automatic temperature balance judgment.
5.Safe, reliable test process with user-friendly operation.
6.Custom-molded refractory furnace with excellent heat insulation performance.
7.High-quality bar heaters enable furnace temperature to rise to 900℃.
8.PLC + NAbview professional software displays real-time curves of in-furnace temperature, sample center temperature, and sample surface temperature, supporting automatic control, data collection/processing, storage, and measurement result output.
9.Top-mounted observation mirror for easy visualization of sample combustion status in the furnace.
10.Installed thermocouples: in-furnace thermocouple, sample center thermocouple, and sample surface thermocouple.
11.360° rotatable and liftable thermocouple scanner.
12.Stainless steel inverted conical air stabilizer and cylinder guide shield.
13.Software automatically controls heating rate and stages furnace body temperature.
14.Software includes current regression criteria for temperature balance judgment.
Related technical parameters:
Furnace specification | D: Φ40cm; High:1.5m |
80 / 20 Nickel Chrome resistance band specifications | Thickness: 0.2mm; width: 3mm |
Sample specification | Φ45×50(mm) |
Heating furnace pipe parameter | D: Φ75mm±1mm; De: Φ90mm±1mm; H:150mm±1mm |
Nickel chromium-nickel silicon armoured thermocouple | De:1.5mm; Silk diameter: 0.3mm |
Environmental conditions | -5℃—40℃ |
Supply voltage | AC220V±10%, 50Hz,Current≤10A |
Temperature stability time of furnace | From room temperature to 750 C≤ for 1h |
The maximum temperature in the furnace | 900℃ |
Temperature in furnace | 750±5°C,drift ≤2℃ within 10min |
Temperature measurement accuracy | ≤±0.5℃ |
Heating power | 800-1000W |
Maximum power | 1.5KW |
Heater voltage | 0~100V adjustable |
Standard:
GB/T5464-2010 GB/T8624 ISO1182 Fire reaction test for building products- - -non-combustible test
![]() |
ISO 1182 220V Combustion Testing Equipment , Building Materials Non Combustible Furnace Images |